Main
page
Production
Dust meter
Set for dioxins
Flowmeter
Meter CMWT-2007Accessories
Type SP-sst
Type SP-sot
Heads
Segments
Tubes
Dust separators for internal
filtration
Dust separators for external
filtration
Set for sampling fum.
Condensation and filtration module
Set for coal dust
Temperature regulator RT-200
Suck-in unit AZ-15
Service and calibration
Training
Measurements
Service heating boilers
Consultations
and sale
E-mail
|
Aspiration speed
probe
SP-sot, SP-sotg, SP-sotgr type
|
The Innovation &
Implementation Firm EMIOR offers an
aspiration speed probe (SP-sot) with a "S"
taken off type head, including an aspiration
route made of titanium. The probe is destined for
sampling purposes for determination of heavy
metal and dioxin presence in the gases being
tested. View of the probe is shown in the drawing
below.
Probe jacket
is made of stainless steel. Special variant of
the probe may include a heating element (SP-sotg).
Then this element is a single-core chrome &
nickel wire in the insulation made of mineralised
titanium oxide. The heating element is laid
directly in the internal titanium aspiration tube
and it is supplied with a safe voltage of 24 V.
Any 230/24 V protective transformer with
capacity not lower than 160 W can be used to
supply power to the heating element. The supply
cable is ended with a plug protected from dust
and moisture.
At request
of the buyer, the probe can be made in a variant
with temperature adjustment up to 200°C (SP-sotgr).
Then a temperature sensor is fitted at the
distance of approx. 0.6 m from the probe end on
the aspiration tube under the probe jacket. The
probe can be used with a temperature regulator,
RT-200 type. Such regulator is offered by EMIOR.
There is
yoke suitable for standard measuring stubs (M64 x
4) fitted on the probe jacket.
| Technical data |
| Probe head type |
"S" taken off |
| Minimal total probe length |
1600 mm |
| Minimal measuring length of
the probe |
1300 mm |
| Maximum probe length |
3000 mm |
| Probe jacket diameter |
38 mm |
| Supply voltage |
24V AC |
| Power taken for supply |
160 VA |
| Temperature sensor type |
PT 100 |
| Outlet connector type |
internal thread
M 22 x 1,5 mm |
| Supply connector protection
degree |
IP66 |
| Length of the connected
supply cable |
3,5 m |
| Accessories |
protective transformer