Microcomputer
flowmeter
DFM 1197
|
A dimensionally small,
portable measuring instrument enjoying a lot of
interest. The flowmeter is manufactured according
to solutions developed in EMIOR. DFM
1197 is based on micro manometer measurement of
pressure differences with its own temperature
probe and moisture measurement probe, measurement
of barometric pressure.
It can be used with any type of impact
pressure (accumulation) probes (tubes), as well
as measurement reducers. It is only necessary to
know their calibration constant values K. It is
applied for measurement of gas rates. Final
results are presented in the form of a report
with calculated values of gas stream referred to
real and stipulated conditions both for dry and
moist gas.
The DFM
1197 flowmeter is applied in measurement of gas
rate using an accumulation tube, possibly with
the option of connecting an anemometric head, TAD
type, to measure low rates. In case the flowmeter
is provided with a probe for measuring relative
moisture, the instrument enables determination of
gas moisture degree. The unit is particularly
useful to identify rate profile by registration
of dynamic pressure distribution in the duct
section being tested. The extended programme of
flowmeter operation covers ongoing calculations
of gas rate in the duct. Final results are shown
in the form of a report including calculated
values of gas stream referred to actual and
stipulated conditions both for dry and wet gas.
One additional feature of the unit is the
possibility of using it as a recorder for
measurement data from gas multi-analyser, MULTOR
610 type, where also rate measurement can be
conducted during the registration taking into
account changes of gas density resulting from
ongoing change of composition, also from the
change of water vapour level in the gas being
tested. In the case of such application of the
unit, the flowmeter registers and processes data
from the analyser and calculates volume of bulk
stream for individual gas compounds based on the
gas rate measurement. During the measurement it
is possible to set up the time of averaging for
the registered values. Memory for measurement
data enables long term registration. For example,
for the averaging time of every minute it is
possible to save data from the period of approx.
24 hours. The programme enables calculation of
concentration for the given reference oxygen
value in dry gas which is consistent with current
legal requirements.
The DFM
1197 flowmeter is addressed to environmental
protection services in heat & power
generating plants, power stations, steelworks and
big industrial sites, and also to testing,
control and environmental protection units. It
can also be applied in laboratory systems in
measurement of flow in closed pipe ducts equipped
with any type of measurement reducer. The
algorithm of the DFM 1197 flowmeter operation is
consistent with recommendations of PN-Z-04030-7:1994
and PN-87/M-34129 standard.
The DFM
1197 flowmeter enables:
- co-operation
with any type of probes for accumulation
pressure measurements;
- automatic
measurement of barometric pressure and
ambient temperature;
- automatic
measurement in the duct of: accumulation
pressure, relative pressure and
temperature of gas, and optionally
relative moisture of gas;
- automatic
compensation of supply voltage
fluctuation and converter zero drift
compensation;
- calculating
results and registering data in the
memory;
- automatic
recording of source values such as:
temperature of the units' environment,
temperature of gas in the duct,
accumulation pressure, relative pressure,
barometric pressure, barometric pressure
and relative moisture, as well as
recording additional information such as:
date and time of starting the
measurement, number of measuring axis,
type of accumulation probe applied,
number of current measurement and
registration on the object in the given
measurement series identified by the
code, and area of the measuring section
in the duct being tested;
- calculating
and recording outcome values including:
gas rate in the selected points of the
measuring section, degree of gas moisture
and, with respect to co-operation with
MULTOR 610 analyser: CO2
proportion calculated based on O2
proportion and the given CO2max
value or type of fuel, density of dry gas
calculated from measured proportions of
individual gas compounds and density of
moist gas, based on measurement of
relative moisture or the given gas
moisture degree;
- automatic
preparation of measurement reports in the
form of a sheet:
- measurement
sheet including subsequently
registered values of individual
source values;
- results
of flow measurements including
calculated value of gas volume
stream, both for real and
stipulated conditions, for wet
and dry gas;
- results
of measuring gas contamination (option
with an analyser) including bulk
streams (emissions) of given gas
compounds;
- review of
results on alpha-numerical display;
- printing
reports on a computer printer or thermal
object printer;
- transmission
of report forms and measurement data to
personal computers in order to save them
on a hard disk or floppy disks and
potentially further reprocessing e.g. in
EMISJA programme.
The DFM 1197
flowmeter unit contains a special computer system
which controls all functions in the measuring and
reporting process. The software used enables
flexible meeting of individual user's
requirements with respect to data processing.
The
following sensors were used in the unit:
differential pressure, relative pressure and
barometric pressure sensor. A temperature probe
is used to measure gas temperature in the duct. A
hygrometric probe (optional) is used to measure
relative moisture of gas. A RS232C connector is
used to send measuring data from the analyser. A
galvanic separation is used on external measuring
routes to prevent influence of electrostatic
interference, and also anti-interference ferric
filters are included reducing impact of
interference coming from radio transmitters /
converters and mobile telephony.

DFM 1197 flowmeter - front view
TECHNICAL
DATA
| Values being
measured |
pressures: differential,
relative and barometric; temperature: gas
and environment, relative moisture (optional)
|
Measuring range for:
differential pressure
gas temperature
relative moisture |
-1250 ÷ +1250 Pa (-625 ÷ 625 Pa -
optional)
0 ÷ 600 °C (0 ÷ 1100 °C - optional)
0 ÷ 100 % RH, tmax=200 °C,
wmax=20 m/s |
| Measuring range of gas rate
(for p ~ 1.2 kg/m3) |
3 ÷ 30 m/s (0-30 m/s -
anemometer) |
| Operating temperature |
5 ÷ 50 °C |
| Acceptable air moisture |
3 ÷ 80 % |
| Outcome values |
gas volume stream, gas rate,
concentration and bulk stream of tested
compounds |
| Method of displaying results
and messages |
liquid crystal display with
the possibility of lighting, 2 lines x 40
characters |
| Method of showing results |
display, printer, computer |
Total measuring error:
pressure
temperatures
relative moisture |
2 %
2 %
2.5 % max. |
| Working time with charged
batteries |
>= 8 hours |
| Dimensions |
225 x 115 x 200 mm |
| Weight |
3 kg |
| Standard outfit |
cover, feeder, belt, cable
RS232, temperature probe (up to 600 °C)
|
| Accessories |
moisture measuring probe,
accumulation tube choice type |
|